176 Rockwell Automation Publication 1441-UM001B-EN-P - September 2012
Chapter 6 Multi-channel Measurements
Set Up Offroute Orbit Measurements
1. Select Data Collection from the Main Menu and press the Enter key.
The Data Collection screen appears.
2. Select the route and measurement point.
See Select a Route on page 131
for more information.
3. Press F2 (Offroute) to collect an Offroute measurement.
The Offroute screen appears.
4. Select the Orbit icon to store an Offroute orbit measurement at the
current point.
The Offroute Orbit Setup screen appears.
5. Select the parameter and press the Right arrow to display the choices.